Beacon Lab System Specifications
|
Performance Specifications
|
Measurement time:
Wavelength range:
Wavelength accuracy:
Wavelength repeatability:
Spectral resolution:
Dynamic range:
Photo-metric accuracy:
Mode of measurement:
Sample cell volume:
Fiber Optic Cable
10 meter standard length |
30 seconds
800-1100nm
<0.015 nm
<0.01 nm
<0.6 nm
1:10,000
<0.1%
transmission
60ml
(can be extended)
|
|
|
Computer Processing
Unit
|
Standard Industrial PC
Operating system MS Windows
|
Physical
Characteristics
|
|
Main Analyzer |
Laboratory Sample System |
Weight
Height
Depth
Width
|
20 kg (44 lb.)
230 mm (9 inches)
660 mm (26 inches)
530 mm (21 inches)
|
24 kg (53 lb.)
720 mm (28 inches)
340 mm (134 inches)
530 mm (21 inches)
|
| Operating Conditions |
Ambient Temperature
Relative humidity
Supply voltage
|
0 C-45C
(32-113F)
30%-90% non-condensing
120/220/240 VAC, 50/60 Hz (3 A max.)
|
|
| Area Classifications |
General Purpose
Specifications are subject to change without notice.
Back to Products | Back to Lab |