Beacon Lab System Specifications

 


Performance Specifications

Measurement time:
Wavelength range:
Wavelength accuracy:
Wavelength repeatability:
Spectral resolution:
Dynamic range:
Photo-metric accuracy:
Mode of measurement:
Sample cell volume:

 

Fiber Optic Cable
10 meter standard length

  30 seconds
  800-1100nm
<0.015 nm
<0.01 nm
<0.6 nm
1:10,000
<0.1%
transmission
60ml



(can be extended)

            

Computer Processing Unit

Standard Industrial PC
Operating system       MS Windows
 

Physical Characteristics

Main Analyzer Laboratory Sample System
Weight
Height
Depth
Width
20 kg (44 lb.)
230 mm (9 inches)
660 mm (26 inches)
530 mm (21 inches)
24 kg (53 lb.)
720 mm (28 inches)
340 mm (134 inches)
530 mm (21 inches)
Operating Conditions
Ambient Temperature
Relative humidity
Supply voltage
0 C-45C (32-113F)
30%-90% non-condensing
120/220/240 VAC, 50/60 Hz (3 A max.)
Area Classifications
General Purpose


Specifications are subject to change without notice.

Back to Products | Back to Lab