Beacon 3000 Specifications |
Field Unit |
Operating
Conditions
|
Ambient Temperature:
Minimum Sample Temperature:
Maximum Sample Temperature:
Sample Conditioning Requirement:
Maximum Inlet Pressure:
Flow Rate Requirement:
Sample Phase
|
-40C to +70C (-40F to 158F)
Application Specific
Above cloud pt. up to 160C (320F)
Haze-free
550 psi ( 40 bars )
1 litre/min to 3 litre/min
Liquid
|
Dimensions
|
Weight:
Diameter:
Depth:
|
Approximately 7 kg (15 lb.)
2.54 cm (1 inches)
38 cm (15 inches)
|
Routine Maintenance
None |
Area Classification
Intrinsically Safe |
Main Analyzer Unit |
Performance Specifications
|
Cycle time: 10-30 sec / stream
Main Analyzer Unit to Field Units maximum distance: 3 km (2 miles)
Multiplexing Capability: up to 8 field units
Outputs:
-
Modbus over RS-485
- Modbus over TCP/IP
- Analog (optional)
|
Operating
Conditions
|
- Ambient Temperature: 0C - 45C (32F - 113F)
- Relative Humidity: 30% - 90% non-condensing
- Power Consumption: 100-240 VAC, 47-63 Hz, 0.63 A Via on-line UPS
|
Dimensions
|
Weight:
Height:
Depth:
Width:
|
Approximately 95 kg (200 lb.)
2000 cm
800 cm
600 cm
|
Routine Maintenance
Replacement of light source every
6 months.
|
Area Classification
General Purpose |
Specifications are subject to change without notice |
|